Understanding A-C-Z noise frequency weightings

Noise Measurement

Understanding A-C-Z noise frequency weightings

By: Callum Wallis


What are the different frequency weightings and why do you need to know?

Understanding A-C-Z noise frequency weightings ad the differences between them is one of the most important things you need to know about when measuring sound. This is because picking the wrong sound level meter weighting on your sound level pressure meter could mean your results become irrelevant for the purpose and at worst invalid (not legally compliant). 

You may have noticed that some sound meters allow you to choose the frequency weighting you want to measure noise at. The three most commonly used decibel weightings are ‘A’, ‘C’ and ‘Z’ as defined in the sound level meter standards IEC 61672:2013 (BS EN 61672-1:2013), but which one do you choose? We explain this is more detail below, but if you own one of our Pulsar Nova sound level meters which measures all of these frequencies simultaneously, you need not ever worry about picking the wrong weighting. 

What are ‘A’, ‘C’ and ‘Z’ weightings

A-weighting – (A-frequency-weighting). A sound level meter weighting that makes its readings conform to a notional human hearing response. It is defined in various international standards such as the IEC 61672, as well as in various national standards such as ANSI S1.4. (USA). ‘A’ Weighted is the most commonly used and covers the full frequency range of 20Hz all the way up to high frequency 20 kHz. The human ear is most sensitive to sound frequencies between 500 Hz and 6 kHz (especially around 4 kHz) whilst at lower and higher frequencies the human ear is not very sensitive. The ‘A’ weighting adjusts the sound pressure level readings to reflect the sensitivity of the human ear and is therefore mandated all over the world for hearing damage risk measurements. Any approved sound level meter meeting IEC 61672 is mandated to incorporate at least an A-weighting filter. Measurements are commonly displayed as dB(A) or dBA or as LAeq, LAFmax, LAE.

C-weighting – (C-frequency-weighting). The C-weighted frequency looks more at the effect of low-frequency sounds on the human ear compared with the A-weighting and is essentially flat or linear between 31.5Hz and 8kHz, the two – 3dB or ‘half power’ points. Peak Sound Pressure Measurements are made using the C- frequency weighting. This is c-weighted peak is for measuring impulse noise and is referred to as CPeak . Measurements are typically displayed as dB(C) or dBC. Or for example as LCeq, LCPeak, LCE – where the C shows the C-weighting.

Z-Weighting – (Z-frequency-weighting). Z-weighted is the flat frequency response of 8Hz to 20kHz (+/- 1.5dB), this is the actual noise that is made with no weighting at all for the human ear (Z for zero).  Often used in octave band analysis and for determining environmental noise. Measurements made are displayed as dB(Z) or dBZ, or LZeq, LZFmax, LZE – where the Z shows the use of the Z-weighting. 

The sound level meter standard IEC 61672 specifies the performance and tolerances for the frequency weighting curves to be used.

Frequency (Hz)631252505001k2k4k8k16k
A-weighting (dB)– 26.2– 16.1– 8.6– 3.201.21-1.1– 6.6
C-weighting (dB)– 0.8– 0.2000– 0.2– 0.8– 3.0– 8.5
Z-weighting (dB)000000000

Why do I need to know this?

The human ear is most sensitive to sound frequencies between 500 Hz and 6 kHz. When measuring sound pressure level variations, especially for potentially damaging noise levels for workplace noise, it is important that the sound level meter is able to give an accurate representation of what the human ear actually hears. Frequency weightings do this by giving more weight to different frequencies over others (i.e. emphasising some frequencies and de-emphasising others). 

A-weighting for noise at work 

When measuring the impact of noise at work on hearing the A-weighted noise measurements should be presented (commonly displayed as dB(A), (correctly written as LAeq)). Representative ‘A’ weighted average noise level readings should be taken for each task undertaken by an employee and then using software or the HSE’s calculator determine an individual’s exposure level. 

Current level and limits dB(A) of an 8-hour period

Current levels of limitsdB(A) / LAeqdB(C) (LCPeak)
Lower Exposure Action Values80 daily or weekly135 daily or weekly
Upper Exposure Action Values85 daily or weekly137 daily or weekly
Legal Limit87 daily or weekly140 daily or weekly

C-weighted decibels for noise at work

People often forget the need to assess the risk from any impulsive noise (very sudden short-lived noises, bangs and crashes). The C-weighting is used for this to give us the peak sound pressure for the impulsive noise that the human ear is exposed to dB(C) (or LCPeak). 

Most modern sound level meters and dosimeters will measure both the LAeq and the LCpeak simultaneously.


Featured Category


Browse Our Range


Featured eBook


Download Now

Related news stories

Noise Assessment Kits for Safety Professionals
Noise Assessment Kits for Safety Professionals

21st Mar 2009

Pulsar Instruments are pleased to announce the introduction of their New Noise Assessment Kits for Safety Professionals that give the Health & Safety professional a custom-made solution for resolving their…

Read full story

Noise Risk Assessments – Measuring Reality
Noise Risk Assessments – Measuring Reality

31st Mar 2009

Effective noise risk assessments require the accurate measurement of an individual workers exposure to noise It is important to capture the ‘reality’ of their daily environment rather than one where…

Read full story

Noise Measurement: Easy dose It!
Noise Measurement: Easy dose It!

24th Jun 2010

Pulsar Instruments Plc have recently launched their new and improved doseBadge system. The noise measurement kit now includes a new ‘fast charge’ function on the updated charger unit, meaning the…

Read full story